ThermoFisher Scientific NEW TVA2020 with GPS and BlueTooth Connectivity
Thermo Scientific

ThermoFisher Scientific NEW TVA2020 with GPS and BlueTooth Connectivity

Pine Item #55377

Easy-to-Own and Operate

The TVA2020 Analyzer is 21% lighter than earlier models and more compact than most FID stand alone instruments. The lightweight and compact design reduces fatigue for true portability and is easier to service in the field. In addition, a variety of options are available such as a basic or enhanced probe, carrying case, and hydrogen refill assembly.

Optional Bluetooth communication transmits concentration data to a handheld device containing the LDAR software database thereby eliminating the need to transfer files post-monitoring and provide a greater access to route information.

The Technology You Expect

Configurable for use in diverse applications including U.S. EPA Method 21 monitoring, site remediation, landfill monitoring, and general area surveys, the TVA2020 Analyzer is equipped with an FID to measure organic compounds with high sensitivity. The FID technology allows for a wide dynamic and linear range that produces stable and repeatable responses. The analyzer can be configured with both FID and PID technology for simultaneous detection and enhanced analytical capabilities. This dual configuration is capable of producing a more rapid reading of organic and inorganic compounds as opposed to a single detector technology and provides more comprehesive gas coverage than comparable size devices.

After performing a primary calibration, the TVA2020 Analyzer can be customized by activating internal logging parameters, uploading a monitoring route, establishing a bluetooth connection, setting alarm levels, and activating response factors.

Features:

Reduced Response Times

Increased user-serviceability

Bluetooth Connectivity

GPS

Quantity must be 1 or more
Email for Quote

Specifications

Technical Specifications

AccuracyFID Instrument–±10% of reading or ±1.0 ppm whichever is greater from 1.0 to 10000 ppm \nPID Instrument–±20 % of reading or ±0.5 ppm which
Components MeasuredMost organic and inorganic compounds
Computer RequirementsUSB port Windows XP operating system or higher
Concentration Measurement RangeFID: 0-30000 ppm (methane) PID: 0-2000 ppm (isobutylene)
Data Logging Averaging PeriodsAuto mode - 1 per second to 1 per 999 minutes user selectable\nVOC or FE Mode - 2 to 30 seconds user selectable
DescriptionTVA2020 Toxic Vapor Analyzer
Flow Rate1L/min. nominal at sample probe inlet
Height (English)11.5 in
Height (Metric)29.2 cm
Hertz50/60
InputsBluetooth USB
Depth (English)4 in.
Depth (Metric)10.2vm
OutputsBluetooth USB
RepeatabilityFID Instrument - 2% at 500 ppm of methane \nPID Instrument - 1% at 100 ppm of isobutylene
Serial InterfaceUSB
Voltage100-230 V
Weight (English)FID only—9.2 pounds\nDual—9.4 pounds
Weight (Metric)FID only— 4.17 kg\nDual— 4.26 kg
Width (English)9 in.
Width (Metric)22.9 cm
Analysis TimeT90 = 3.5 seconds
Battery LifeThe battery operating time is 10 hours minimum at 0 °C (32 °F). Fully charged in less than 10 hours. - See more at: https://www.thermoscient
Battery TypeLithium ion
Calibration KitPart Number for FID: CR009UY \nPart Number for dual detector: CR012UH
Detector TypeFID or PID/FID
Fuel99.99% H2 for FID
Min. Detectable LimitThe minimum detectable level is defined as seven times the standard deviation of peak-to-peak noise.\nPID Instrument - 0.5 ppm of isobutylene
Portable Operation TimeMinimum 10 hours
Response TimePID Instrument - Less than 3.5 seconds for 90% of final value using 500 ppm of isobutylene \nFID Instrument - Less than 3.5 seconds for 90%
Temperature (English) Range14° F - 113° F
Temperature (Metric) Range-10°C - +45°C
WarrantyOne year from date of shipment