Dwyer Series 2000 Magnehelic Differential Pressure Gage
Multiple Pressure Ranges Available
Select the Magnehelic ® gage for high accuracy--guaranteed within 2% of full scale--and for the wide choice of models available to suit your needs precisely. Using Dwyer's simple, frictionless Magnehelic ® gage movement, it quickly indicates low air or non-corrosive gas pressures--either positive, negative (vacuum) or differential. The design resists shock, vibration and over-pressures. No manometer fluid to evaporate, freeze or cause toxic or leveling problems.
The Magnehelic ® gage is the industry standard to measure fan and blower pressures, filter resistance, air velocity, furnace draft, pressure drop across orifice plates, liquid levels with bubbler systems and pressures in fluid amplifier or fluidic systems. It also checks gas-air ratio controls and automatic valves, and monitors blood and respiratory pressures in medical care equipment.
Note: May be used with Hydrogen. Order a Buna-N diaphragm. Pressures must be less than 35 psi.
Features:
Bezel provides flange for flush mounting in panel.
Clear plastic face is highly resistant to breakage. Provides undistorted viewing of pointer and scale.
Precision litho-printed scale is accurate and easy to read.
Red tipped pointer of heat treated aluminum tubing is easy to see. It is rigidly mounted on the helix shaft.
Pointer stops of molded rubber prevent pointer over-travel without damage.
Jeweled bearings are shock-resistant mounted; provide virtually friction-free motion for helix. Motion damped with high viscosity silicone fluid.
Zero adjustment screw is conveniently located in the plastic cover and is accessible without removing the cover. O-ring seal provides pressure tightness.
Helix is precision made from an alloy of high magnetic permeability. Mounted in jeweled bearings it turns freely following the magnetic field to move the pointer across the scale.
Calibrated range spring is flat spring steel. Small amplitude of motion assures consistency and long life. It reacts to pressure on diaphragm. Live length adjustable for calibration.
Silicone rubber diaphragm with integrally molded O-ring is supported by front and rear plates. It is locked and sealed in position with a sealing plate and retaining ring. Diaphragm motion is restricted to prevent damage due to overpressures.
Die cast aluminum case is precision made and iridite-dipped to withstand 168 hour salt spray corrosion test. Exterior finished in baked dark gray hammerloid. One case size is used for all standard pressure options and for both surface and flush mounting.
OVERPRESSURE PROTECTION
O-ring seal for cover assures pressure integrity of case.
Samarium Cobalt magnet mounted at one end of range spring rotates helix without mechanical linkages.
Applications:
Bezel provides flange for flush mounting in panel.
Clear plastic face is highly resistant to breakage. Provides undistorted viewing of pointer and scale.
Precision litho-printed scale is accurate and easy to read.
Red tipped pointer of heat treated aluminum tubing is easy to see. It is rigidly mounted on the helix shaft.
Pointer stops of molded rubber prevent pointer over-travel without damage.
Jeweled bearings are shock-resistant mounted; provide virtually friction-free motion for helix. Motion damped with high viscosity silicone fluid.
Zero adjustment screw is conveniently located in the plastic cover and is accessible without removing the cover. O-ring seal provides pressure tightness.
Helix is precision made from an alloy of high magnetic permeability. Mounted in jeweled bearings it turns freely following the magnetic field to move the pointer across the scale.
Calibrated range spring is flat spring steel. Small amplitude of motion assures consistency and long life. It reacts to pressure on diaphragm. Live length adjustable for calibration.
Silicone rubber diaphragm with integrally molded O-ring is supported by front and rear plates. It is locked and sealed in position with a sealing plate and retaining ring. Diaphragm motion is restricted to prevent damage due to overpressures.
Die cast aluminum case is precision made and iridite-dipped to withstand 168 hour salt spray corrosion test. Exterior finished in baked dark gray hammerloid. One case size is used for all standard pressure options and for both surface and flush mounting.
OVERPRESSURE PROTECTION
O-ring seal for cover assures pressure integrity of case.
Samarium Cobalt magnet mounted at one end of range spring rotates helix without mechanical linkages.
Service | Air and non-combustible compatible gases. (Natural Gas option available.) |
Note | May be used with hydrogen. Order a Buna-N diaphragm. Pressures must be less than 35 psi. |
Wetted Materials | Consult factory. |
Housing | Die cast aluminum case and bezel with acrylic cover Exterior finish is coated gray to withstand 168 hour salt spray corrosion test. |
Accuracy | ±2% of FS (±3% on -0 -100PA -125PA -10MM and ±4% on -00 -60PA -6MM ranges) throughout range at 70°F (21.1°C). |
Pressure Limits | -20 in Hg. to 15 psig. (-0.677 bar to 1.034 bar); MP option; 35 psig (2.41 bar) HP option; 80 psig (5.52 bar). |
Overpressure | Relief plug opens at approximately 25 psig (1.72 bar) standard gages only |
Temperature Limits | 20 to 140°F.* (-6.67 to 60°C). |
Size | 4\" |
Mounting Orientation | Diaphragm in vertical position. Consult factory for other position orientations. |
Process Connections | 1/8\" female NPT duplicate high and low pressure taps - one pair side and one pair back. |
Agency Approval | RoHS. (Note: -SP models not RoHS approved.) |