VIG 210 METHANE/ETHANE
VIG

VIG 210 METHANE/ETHANE

Pine Item #54393

Now available for rent, VIG model-210 is a microprocessor based, oven heated methane / ethane / non-methane-ethane / total hydrocarbon gas analyzer designed for high accuracy, sensitivity and stability. 

The Model-210 uses two independent flame ionization detectors (FIDs), one to measure total hydrocarbons and the second coupled with a GC Column for the separation of the methane, ethane and non-methane-ethane components.

A sample is fed to the analyzer via an internal heated pump to the first FID for a real time total hydrocarbon reading.  A portion of the sample is trapped and pushed through a column to separate the methane and ethane components and then to the second FID.  Any remaining sample in the column is back flushed through the column to obtain the non-methane-ethane components.

All components that come in contact with the sample through analysis are maintained in a temperature-controlled oven to prevent condensation, and to provide repeatable, reliable performance in the analysis of a wide variety of hydrocarbon concentrations in gaseous mixtures or in ambient air.

Features:

Adjustable alarm and oven settings

Adjustable alarm and oven settings

Heated sample pump heads

Teflon isolated detectors (FIDs)

Two stage sample filter with exchangeable sintered stainless steel elements

Applications:

Adjustable alarm and oven settings

Adjustable alarm and oven settings

Heated sample pump heads

Teflon isolated detectors (FIDs)

Two stage sample filter with exchangeable sintered stainless steel elements

Quantity must be 1 or more
Email for Quote

Specifications

Technical Specifications

Measuring Method2 Oven Heated Flame Ionization Detectors (FIDs)
Zero & Span NoiseLess than 0.2% of full scale
Zero & Span Drift+/- 1% full scale per 24 hours
Linearity/Repeatability/StabilityWithin 1% of full scale through all ranges
Oxygen SynergismWithin 1% of full scale within selected range
Response TimeTotal-Within 5 seconds to 90% of final reading (Continuous real time reading)
NULLEthane-Approximately 60 seconds updated every 4 minutes
Ambient TermperatureFrom 50°F to 120°F
Flow Rate4 Liters/Minute (standard) or 10 Liters/Minute (Upon request)

Dimension

USMetric
Dimensions19 x 16.75 x 24 in-

Weight

USMetric
Weight55 lbs-