Thermo Scientific Model 146i Multi-Gas Calibrator
Thermo Scientific

Thermo Scientific Model 146i Multi-Gas Calibrator

Pine Item #54946

The Thermo Scientific Model 146i Multi-Gas Calibrator supplies precise levels of ozone, carbon monoxide, non-methane hydrocarbons, sulfur dioxide, nitric oxide, and nitrogen dioxide or other gases that a user may require. The gas levels are used to calibrate instruments that perform zero, precision and level 1 span checks, audits and multipoint measurements. The design of the Thermo Scientific Model 146i calibrator meets or exceeds all published U.S. EPA requirements for multipoint calibration, audit, Level 1 and 2 span and precision checks. Options include Gas Phase Titration, UV Photometer, and Permeation Oven. The mass flow controller, ozone generator, permeation tube oven, power supply, and solenoid valves are integrated into a single microprocessor-controlled unit. This permits easy to use, menu-driven software and a consistent set of operation screens for all calibration procedures. Additionally, the microprocessor can make many of the necessary calculations, thereby freeing the operator from having to make the calculations in
the field or laboratory. If desired, the Model 146i calibrator can be operated remotely by a data logger to perform multipoint calibrations.

Features:

Ethernet connectivity for effi cient remote access

Enhanced user interface with one button programming and large display screen

Flash memory for increased data storage and user downloadable software

Enhanced electronics design optimizes product commonality

Gas Phase Titration for generation of O3 and NO2

Quantity must be 1 or more
Email for Quote

Specifications

Technical Specifications

Flow Measurement Accuracy+/- 2% of reading or 1% of full scale whichever is less (20 to 100% full-scale)
Linearity of Mass Flow Measurement+/- 0.5% of full scale
Repeatability of Mass Flow Measurement+/- 2% of reading or 1% of full scale whichever is less (20 to 100% full-scale)
Flow Range of Dilution Air0-10 SLPM. Optional ranges: 0-5/0-20 SLPM
Flow Range of Cylinder Gases0-100SCCM. Optional ranges: 0-50/0-200 SCCM
Zero Air Requirements10 SLPM @ 30 PSI Optional ranges: 20 SLPM @ 30 PSI
Calibration Gas Input Ports3 optional 6
Dilutent Gas Input Ports1
Response Time99% of target (at manifold) within 60 seconds
Ozone Generator OptionMaximum output: 1 ppm @ 6 SLPM. Minimum output: 10 ppb @ 6 SLPM
Photometer SystemFull Scale Range: 100 ppb to 5 ppm user selectable. Linearity: 1% of Full Scale.
Precision:1 ppb. Response Time: 180 Seconds to 95% of Target. Minimum Detectable Limit 3 ppb.
Test Channel Analog6 @ +/- 100 mv 1510 volts (user selectable)
Digital Control Outputs10 relay and 8 24vdc solenoid
Digital Control Inputs16
Temperature Range0°- 45° C
Weight51 lbs (58 lbs for 220-240 VAC)
Dimensions16.75” W x 8.62” H x 23” D
Power100 VAC 50/60Hz 115 VAC 50/60Hz 220-240 VAC 50/60Hz 275 watts (with all options)