The new Fluke 983 Particle Counter is the preferred choice for HVAC and IAQ professionals. From filter testing to IAQ investigations, the Fluke 983 is the portable solution for determining airborne particle concentrations. Use the Fluke 983 to immediately respond to occupant complaints, or as part of a comprehensive preventive maintenance program.
With the Fluke 983, you can:
|
With expanded data logging and six-channel particle size display, the Fluke 983 allows users to run more tests quickly, with less time spent cycling through screens to obtain the data. The Fluke 983 is lightweight and easy to use in any position, with a comfortable hand strap and rugged rubber holster.
The Fluke 983 Particle Counter simultaneously measures and displays six channels of particle size distribution, temperature, and humidity. This compact, lightweight, self-contained tool allows for one-handed operation. The Fluke 983 maintains its accuracy when used in any position. The 8-hour rechargeable battery together with a large 5000 sample record capacity allows for uninterrupted performance during air quality surveys. Fluke 983 is the ideal tool for determining the size distribution of airborne particles or tracking down a particle source.
|
Features:
Compact and lightweight
Graphic liquid crystal display with backlight
AC adapter
Rechargeable battery
Simple user interface
Applications:
Monitoring cleanrooms
Indoor air quality investigations
Monitoring gowning rooms
Testing filter seals
Locating particle contamination sources
Monitoring particle size distributions
Channels | 0.3, 0.5, 1.0, 2.0, 5.0, 10.0 µm |
Flow Rate | 0.1 cfm (2.83 L/min) controlled by internal pump |
Count Modes | Concentration, totalize, audio |
Counting Efficiency | 50% @0.3 µm; 100% for particles > 0.45 µm (per JIS B9921:1997) |
Zero Count | 1 count/5 minute (JIS B9921:1997) |
Coincidence Loss | 5% at 2,000,000 particles per ft3 |
Relative Humidity | ± 7%, 20% to 90% non-condensing |
Temperature | ± 3 ˚C, 10 ˚C to 40 ˚C (50 ˚F to 104 ˚F) |
Data Storage | 5000 sample records (rotating buffer) of date, time, counts, relative humidity, temperature, sample volumes, alarms, label |
Alarms | Counts, low battery, sensor fail |
Dimensions | US: 4.5 x 8.25 x 2.25 in. Metric: 11.43 x 20.96 x 5.72 cm |
Weight | US: 2.2 lb. Metric: 1.0 kg |