The HI 9125N heavy-duty waterproof pH/mV meter has been designed with
the latest microprocessor technology to provide laboratory results and
accuracy under harsh industrial conditions.
A large dual-level LCD displays the pH and temperature
simultaneously. Graphic symbols are provided for helping the user during
the calibration process.
The pH calibration procedure is automatic with 5 memorized buffer
values (4.01, 6.86, 7.01, 9.18 and 10.01), buffer recognition to avoid
errors and automatic temperature compensation.
The HI 9125N can be used with ORP (Oxidation Reduction Potential)
electrodes. The resolution automatically switch from 0.1 to 1 mV when
the reading exceeds 400 mV.
The meter is also provided with the BEPS (Battery Error Prevention
System) feature, which automatically turns the instrument off when the
battery level is too low to ensure reliable readings.
Slim Design
Flat Back for Better Grip
New Larger LCD
Tutorial messages displayed
Electrode Diagnostics
Order Information:
HI 9125N is supplied with HI 1230B combination double-junction,
gel-filled pH electrode, HI 7669/2W stainless steel temperature probe
with 1 m (3.3') cable, pH 4.01 & pH 7.01 buffer solutions (20 mL
sachets), 100 mL plastic beaker, instruction manual and rugged carrying
case.
Specifications
Range
pH
-2.00 to 16.00 pH
Range
mV
±699.9 mV; ±1999 mV
Range
Temperature
20.0 to 120.0°C / -4.0°F to 248.0°F
Resolution
pH
0.01 pH
Resolution
mV
0.1 mV; 1 mV
Resolution
Temperature
0.1 °C / 0.1 °F
Accuracy
pH
±0.01 pH
Accuracy
mV
±0.2 mV; ±1 mV
Accuracy
Temperature
±0.4 °C / ±0.8 °F
pH Calibration
automatic 1 or 2 point with 5 memorized buffer values (4.01, 6.86, 7.01, 9.18, 10.01)
Offset Calibration
±1 pH
Slope Calibration
From 80 to 108%
Temperature Compensation
automatic or manual, -20 to 120°C (-4 to 248°F)
pH Electrode
HI 1230B, double junction, gel filled, BNC connector and 1 m (3.3’) cable (included)
Temperature Probe
HI 7662 with 1 m (3.3’) cable (included)
Input Impedance
1012 Ohm
Battery Type / Life
(3) 1.5V AAA / approx 200 hours of continuous use (with backlight off); auto-off after 20 minutes of non-use (can be disabled)