24/7 Technical Support
Drager-X-act 5000 Get Quote
 
Question?

 

 

The automatic tube pump Dräger X-act® 5000 is the first all-in-one solution designed for measurements with Dräger Short-term Tubes and for Sampling Tubes and Systems. Ease of operation and a high degree of reliability compliment the measurement and sampling of gases, vapors and aerosols.

Key Features

  • Measurement in Technical Gases
  • Automatic Transfer of Measurement Parameters
  • Simplicity of Operation
  • Direct Settings for Sampling
  • New Pump Concept

Technical Specifications

Short-Term Measurements using Dräger short-term tubes\nNumber of strokes: adjustable, 1 to 199 strokes\nStroke volume: 100 mL
Sample Taking Flow rate range: 0.1 to 2.0 L/min\nResolution 0.1 to 1.0 L/min: 0.1 L/min ± 5 %\nResolution 1.0 to 2.0 L/min: 0.2 L/min ± 5 %\n\nSampling time:
Display two parts: segment and matrix
Menu Languages Danish, Dutch, English, Finnish, French, German, Italian, Norwegian, Polish, Spanish, Swedish
Use of Extension Hose up to 30 meter (98 ft.)
Dimensions (H x W x D) approx. 175 x 230 x 108 mm (7“ x 9“ x 4.25“)
Weight (without battery pack) approx. 1.6 kg (3.5 lbs)
Temperature during Storage -20 to 55 °C (-4 to 131 °F)
Temperature during Operation 5 to 40 °C (41 to 104 °F)
Humidity 0 to 95% r. h., non-condensing
Pressure 700 to 1300 hPa (10.2 to 18.9 psi)
Rechargeable Battery NiMH, T4 7.2 V, 1500 mAh (charging time < 4 hours)
Alkaline Battery Pack, T4 6 AA cells, (see battery pack label for approved battery types)
Battery capacity at 25 °C (77 °F) Dräger short-term tubes: >1,000 strokes\nSample taking: up to 12 hours, depending on adjusted flow rate
ATEX Ex ia IIC T4 Ga\nEx ia I Ma\nI M 1 / II 1G
MED Steering Wheel approval 96/98 EC
UL Class I, Div. 1, Group A, B, C, D\nClass II, Div. 1, Group F, G\n5 °C = Ta = + 40 °C Temp. Code T4 Ex ia
cUL Class I, Div. 1, Group A, B, C, D\nClass II, Div. 1, Group F, G\n5 °C = Ta = + 40 °C Temp. Code T4 Ex ia
IECEx Ex ia IIC T4 Ga
CE-mark according to 2004/108/EC and 94/9/EC