24/7 Technical Support
MDA TLD-1 Monitor Honeywell Analytics Get Quote
 
Question?

 

 

The ChemKey TLD monitor is ideal for temporary facility monitoring and R&D projects, as well as in spill remediation and emergency response. The unique technology of the instrument uses a chemically impregnated tape that changes color in direct proportion to the amount of gas present, even at parts per billion. The tape provides a direct read of the gas and ignores non-target gases. CHEMKEYS AVAILABLE: (HF) (HCL) (H2S04) (HNO3)

Key Features

  • Direct reading of substance-specific exposure
  • Rechargeable battery provides up to eight hours of continuous monitoring.
  • The LCD readout displays actual gas concentration in ppm or ppb.

Technical Specifications

Detectable Gases Amines, diisocyanates, hydrazines, hydrides, hydrogen cyanide, hydrogen sulfide, mineral acids, oxidizers, phosgene, sulfur dioxide
Response Time As fast as 10 seconds, gas dependent
Battery Rechargeable lead-acid battery 8 hours operation, or 115V
Chemkey's Available HF 6PPM, HCL 5PPM, H2SO4 250PPB, HNO3 2 PPM
Chemkey's Available HF 6PPM, HCL 5PPM, H2SO4 250PPB, HNO3 2 PPM

Dimensions

  US Metric
H x W x D 6.5 x 8.4\" x 7.0\"\" 165 x 213 x 178 mm

Weight

  US Metric
H x W x D 9 lbs 4 kg

Consumables

Tubing
Tubing