The PQ100 is the only truly portable PM10 sampling unit on the market
today. With microprocessor control plus mass flow control,
programmability and download/storage capacity, this advanced system of
pump and accessories are designed for any sampling application that
requires portability plus regulatory compliance.
Because of the mass flow sensor, no flow corrections are
needed. The units are supplied with full calibration to engineering
standard conditions at the usual flow rates of 2, 10 and 16.7 lpm.
Additional calibration data (up to 240 sets of conditions) can be stored
for instant access. Calibration is easily confirmed or altered in the
field using a bubble flow meter. The unit display gives elapsed time,
total sample volume, battery state and instantaneous flow rate. The
built-in real time clock and associated microprocessor allows future run
times to be entered.
This microprocessor-controlled air sampling pump is
designated specifically for PM10 sampling. The unit will operate from
its internal battery for 24 hours, but an auxiliary, easily
interchangeable battery greatly extends run time. The built in software
permits programming of start and run time as required. Air flow is
rigidly controlled to ±2% precision with state-of-the art mass flow
sensor. All relevant operations information such as date, time, flow
rate and total flow are displayed. An RS232 port affords downloading of
information plus remote access. During the data transfer, relevant data
such as station number, job code and operator name may be entered. After
filter analysis, the filter weight change may be entered and the system
automatically calculates the air concentration corrected to EPA
standard conditions. The real time clock permits sample start at any
future time.
Applications
Approved by EPA for PM10 sampling (Designation No. RFPS-1298-124)
Suitable for Lead, Asbestos, Oil Mists, Indoor Air Pollutants and Allergens
Offers well Regulated Air Flow (Mass Flow Control) in the 2 to 25 LPM range
Features
Accurate EPA Specified Flow Rate to ±0.5%
Precision Flow to ±2%
Download Software Included
Complete System is Portable and easily usable Indoors or Out
Weatherproof Enclosure Included
24 Hour Sample on Internal Battery
Universal PQ/TSP Sampling Inlet
The publication of an extensive multi year European aerosol
study* of inlet performance confirms the utility of the US EPA louvered
inlet. This is essentially the top half of the McFarland Texas A&M
Federal Reference Method (FRM) inlet without the impactor and water
excluder sections. The Inlet (P/N PQTSP) should be utilized with a
minimum of a 19.5” long inlet sample tube (P/N L20) to avoid bluff body
effects.
* L Kenny, et al. “Aspiration and sampling efficiencies of the TSP and Louvered particulate matter inlets”, J. Environ. Monit, 2005, 7 481-487.
Calibration
When using a specific filter and flow rate (i.e. 47mm Teflo
at 16.7 lpm) this calibration will be maintained in memory indefinitely.
However, calibration should be checked on a yearly basis. Two hundred
and forty calibration locations (from 2.0 to 25.0 in 0.1 lpm increments)
can be programmed into this unit. When you calibrate the above flow
rate/filter, and should you choose to correct for engineering standard
conditions (760mm Hg at 20°C), then every sample you run thereafter at
16.7 lpm using the 47mm Teflo will also remain constant.
Rack/Support Assembly
This assembly is designed to be rigid and easier to use
since the instrument panel is supported at waist height. The legs use
the same "snap and lock" assembly developed for the PQ200 (and are, in
fact, interchangeable). A large water trap has been included in the new
specification for deployment of the instrument in extreme climates.
Power Input / RS 232 Adapter
This device allows all models of the BGI PQ 100 Air Sampler
to be Charged, run on line current coupled to a computer at the same
time. This device is applicable to any PQ 100 ever made but is of
particular use with models having a Version Number higher than 6.0
Downloading
The data from the PQ100 may be downloaded to any IBM type
computer system (5.25 and 3.5 floppy disks are provided). Any IBM
compatible PC running MS-DOS 2.1 or higher with 1 floppy drive and at
least 512K of memory is usable. One serial communication port is also
required. The HP100LX palmtop computer is also usable with our accessory
memory card and adaptor cable. The software also includes a corrections
program which simplifies calculating the flow rate to EPA standards.
After filter analysis, the downloaded information may be reaccessed and
the program will automatically calculate the concentration to EPA
standard conditions of 760mm of Hg and 25°C.
Solar Panel
The BGI Solar Panel accessory for the PQ200 will be extend
its operational time range from 24 hours to an indefinite period
depending upon the amount of ambient sunlight available in the region of
use. As a practical matter the panel with the ballast battery will
provide a minimum of 10-14 days of operation with only minimum ambient
light. When interval sampling on three or six day cycles is contemplated
continuous operations far from a source of power are achievable.
Specifications
PQ100 Sampling Pump
Dimensions
10 x 6 x 9.7 in (25.4 x 15.2 x 24.6 cm)
Weight
19 lbs (8.63 kg)
BGI 167 R
Height
79 in (2 m)
Weight
34 lb (15 kg)
BGI 16.7 PM10 Inlet Kit: Inlet, Tripod, and Filter Holder